Jenway 67 Series Spectrophotometer

E37-31 

   

SPECTROPHO

TOMETERS

SPECTROPHOTOMETERS

All methods in the 67 Series offer comprehensive modes of operation as follows:
Photometrics for simple Absorbance and %T measurements at selected wavelengths.
Kinetics for timed measurement and enzyme reaction monitoring.
Mulit-wavelength for Absorbance and %T measurements at 4 discreet wavelengths.
Concentration and Quantitation for concentration measurements based on a standard, 
factor or calibration curve.

Scanning at an impressive 1500nm/minute, even with data collection at 0.1nm intervals. 
This enables zooming to higher resolutions without the need to re-scan and full post-scan 
re-scaling of either axis.

Model 6700 covers the visible wavelength range with a spectral bandwidth of 4nm.
Model 6705 covers the UV/visible wavelength range with a spectral bandwidth of 4nm.
Model 6715 covers the UV/visible wavelength range with a spectral bandwidth of 1.5nm. 

 



Colour (TFT) LCD with Touch-Screen Interface

 



Sealed QWheel™ Control

 



Save to SD Card

 



Plug-In Accessory Modules

 



Secure Multi-User Operation

 



PC Software Included

 



3 year warranty 
 
 

 

Specifications

6700

6705

6715

Light Source

Tungsten Halogen

Xenon

Xenon

Optics

Sealed, Coated, Split Beam optics

Spectral Bandwidth, nm

4

4

1.5

Stray Light

0.05% at 340nm

0.05% at 220nm

0.05% at 220nm

Wavelength Range, nm

320 to 1100

190 to 1100

190 to 1100

Wavelength Resolution, nm 0.1

0.1

0.1

Wavelength Accuracy, nm

±1.0

±1.0

±1.0

No of Standards

20 with up to 5 replicates of each

Curve Fit Algorithms

Linear, Quadratic and Cubic Functions

Kinetics Time Limits

0 to 999 Secs

Scan Speed

1500nm/minute at 0.1nm data steps

No of Users

10 + Supervisor

No of Methods (Internal 
memory)

1000 with 128mb memory

Results Storage (internal 
memory)

1000 with 128mb memory

Removeable Media

SD Memory Card

Interface

USB, Centronics, Analogue

Code

Description

SPE3500

6700, single cell holder

SPE3502

6705, single cell holder

SPE3506

6715, single cell holder

Accessories

   

Code

Description

SPE3510

Jenway sample chamber with automatic eight cell changer module

SPE3510

Sample chamber with automatic eight cell changer module

SPE3512

Additional carrousel for automatic eight cell changer (SPE3510)

SPE3514

10 x 10mm path length cuvette holder

SPE3516

Test Tube holder for 10mm square cuvettes and 16 and 24mm dia. 
Tubes

SPE3518

Adjustable path length cuvette holder 10 to 100mm

SPE3520

Micro cuvette holder with adjustable aperture

Code

Description

SPE3524

Water heated sample chamber with automatic six cell changer module

SPE3526

Additional carousel for water heated automatic eight cell changer 
(SPE3524)

SPE3528

Sample chamber with sipper module

SPE3530

Sample chamber with Peltier module

SPE3532

Sample chamber with Peltier sipper module

SPE3534

Single cell sample chamber with water heated cuvette holder